Does an IC link error occur at the load step or as the package warms? Use the SANWOOD Temperature Forcing System to investigate intermittent faults under load.
A processor board passes its room-temperature check. Its communication link stays up at idle, yet the receiver occasionally flags a bad packet after the processor switches to sustained traffic. For a semiconductor testing lab, the first question is one of timing. Does the error occur as the electrical load changes, or does the link run cleanly until the package warms? In this IC failure analysis example, the SANWOOD Temperature Forcing System lets the engineer vary local temperature around the powered device while external instruments record link behavior and supply voltage.
Start with the record that prompted the investigation. A final count of bad packets tells the engineer how many the receiver detected during the run. It offers little guidance on which part of a supply waveform to examine. If software polls an error counter, the first increase places the event somewhere between two reads. A receiver timestamp tied to the failed packet can narrow that interval. Check how the receiver clock relates to the electrical capture before using either record to establish which event came first.
Repeat the idle-to-traffic transition on the same board with the firmware, link speed, supply settings and traffic pattern fixed. Mark the start of traffic in the electrical capture and use the same event to align the temperature log. Save a passing transition alongside a failing one. If the supply voltage dips on both, compare the depth, duration and timing of the two dips before attributing the link error to either one.
In Air Mode, the SANWOOD Temperature Forcing System controls the temperature of the supplied air. Attach a thermocouple at a documented point on the IC package and keep the thermal head and probe leads in place. The air setpoint can stay constant while the device dissipates more power under traffic. The package trace then shows how the device surface responds during the workload that produced the original symptom.
Read the records from the traffic marker forward. If the first error falls within the load transition, compare the rail voltage near the device on passing and failing runs. If the receiver handles the initial packets successfully and errors recur later near a similar package reading, use that reading to select temperatures for a follow-up trial. Allow for the response time of the surface sensor. The package trace can look flat during a brief event even as the temperature inside the device changes.
By the time a late error appears, the package has had time to warm and the receiver has processed more packets. Even with a constant error probability per packet, a longer burst gives the receiver more opportunities to detect an error. Give each temperature condition the same packet count, traffic pattern and data rate. Keep the idle interval between bursts consistent so the board has a comparable recovery period before each transition.
In DUT Mode, the system uses temperature feedback from a selected point on the device under test . The SANWOOD system accepts external Type T and Type K thermocouples. Choose setpoints within the device's permitted operating conditions. At each setpoint, let the measured package temperature settle within the tolerance defined for the trial before starting traffic. Keep DUT Mode active and record the package temperature throughout the burst.
That last trace matters when interpreting the result. A warmer setpoint can change both the starting temperature and the temperature profile during transmission. If errors appear sooner, examine the readings leading up to each error before deciding what to test next. The thermocouple measures its attachment point on the package. Estimating junction temperature requires a separate measurement or a suitable thermal model.
For each setpoint, report how many transitions produced at least one error out of the total attempted. On affected transitions, also record the delay from traffic onset to the first detected error. Ten bad packets in one burst and one bad packet in each of ten bursts produce the same final count. The distinction matters to the engineer trying to reproduce the symptom on the next run.
| Observation across repeated runs | Next check |
| Errors cluster near the load step at several DUT setpoints. | Compare rail captures from passing and failing transitions. Check whether the receiver's timing resolution can establish which event came first. |
| Errors appear sooner at the warmer DUT setpoint. | Compare the measured package-temperature traces throughout each burst. Repeat the setpoints in reverse order to see whether the pattern follows temperature or run order. |
| The symptom changes when the thermal head moves. | Restore the original head position and probe routing. Check which neighboring components receive airflow in each position. |
| A reset restores the link while the package is cooling. | Where the test permits, try a reset at a stable package temperature. On a separate run, cool the package without resetting the link. |
An environmental test chamber is useful when the whole board must reach a particular temperature to reproduce the symptom. Keep that result as a reference when applying local airflow to individual regions. If the failure occurs on the first power-up after cold exposure, the order of cooling and power application matters. SANWOOD's guide to investigating IC cold-start failures covers that sequence.
Run a comparable board through the same fixture, traffic sequence and number of transitions. If both boards fail at the same point, examine shared elements such as the supply, fixture or traffic generator. If only the returned board fails, keep the passing board's record with its attempt count. A colleague reviewing a clean trace will want to know how many opportunities the link had to fail.
Give SANWOOD the target temperature at the powered package, the allowed deviation at the sensor and the time available before traffic begins. Include idle and active power dissipation, package dimensions and a fixture drawing showing the thermocouple, supply probe and clearance around the thermal head. If the trial includes cold operation, describe how moisture is controlled around exposed surfaces.
For a configuration review, send the fixture drawing and a short description of the fault sequence to info@sanwood.com. Those details help SANWOOD assess whether the planned airflow can reach the target package while leaving the probes in position to capture the first error.